Failure of a novel silicone-polyurethane copolymer (Optim™) to prevent implantable cardioverter-defibrillator lead insulation abrasions

一种新型硅酮-聚氨酯共聚物(Optim™)未能防止植入式心脏复律除颤器导线绝缘层磨损

阅读:1

Abstract

AIM: The purpose of this study was to determine if Optim™, a unique copolymer of silicone and polyurethane, protects Riata ST Optim and Durata implantable cardioverter-defibrillator (ICD) leads (SJM, St Jude Medical Inc., Sylmar, CA, USA) from abrasions that cause lead failure. METHODS AND RESULTS: We searched the US Food and Drug Administration's (FDA's) Manufacturers and User Device Experience (MAUDE) database on 13 April 2012 using the simple search terms 'Riata ST Optim™ abrasion analysis' and 'Durata abrasion analysis'. Lead implant time was estimated by subtracting 3 months from the reported lead age. The MAUDE search returned 15 reports for Riata ST Optim™ and 37 reports for Durata leads, which were submitted by SJM based on its analyses of returned leads for clinical events that occurred between December 2007 and January 2012. Riata ST Optim™ leads had been implanted 29.1 ± 11.7 months. Eight of 15 leads had can abrasions and three abrasions were caused by friction with another device, most likely another lead. Four of these abrasions resulted in high-voltage failures and one death. One failure was caused by an internal insulation defect. Durata leads had been implanted 22.2 ± 10.6 months. Twelve Durata leads had can abrasions, and six leads had abrasions caused by friction with another device. Of these 18 can and other device abrasions, 13 (72%) had electrical abnormalities. Low impedances identified three internal insulation abrasions. CONCLUSIONS: Riata ST Optim™ and Durata ICD leads have failed due to insulation abrasions. Optim™ did not prevent these abrasions, which developed ≤ 4 years after implant. Studies are needed to determine the incidence of these failures and their clinical implications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。