In vivo analysis of the origin and characteristics of gaseous microemboli during catheter-mediated irreversible electroporation

导管介导的不可逆电穿孔过程中气体微栓子的来源和特征的体内分析

阅读:2

Abstract

AIMS: Irreversible electroporation (IRE) ablation is a non-thermal ablation method based on the application of direct current between a multi-electrode catheter and skin electrode. The delivery of current through blood leads to electrolysis. Some studies suggest that gaseous (micro)emboli might be associated with myocardial damage and/or (a)symptomatic cerebral ischaemic events. The aim of this study was to compare the amount of gas generated during IRE ablation and during radiofrequency (RF) ablation. METHODS AND RESULTS: In six 60-75 kg pigs, an extracorporeal femoral shunt was outfitted with a bubble-counter to detect the size and total volume of gas bubbles. Anodal and cathodal 200 J IRE applications were delivered in the left atrium (LA) using a 14-electrode circular catheter. The 30 and 60 s 40 W RF point-by-point ablations were performed. Using transoesophageal echocardiography (TOE), gas formation was visualized. Average gas volumes were 0.6 ± 0.6 and 56.9 ± 19.1 μL (P < 0.01) for each anodal and cathodal IRE application, respectively. Also, qualitative TOE imaging showed significantly less LA bubble contrast with anodal than with cathodal applications. Radiofrequency ablations produced 1.7 ± 2.9 and 6.7 ± 7.4 μL of gas, for 30 and 60 s ablation time, respectively. CONCLUSION: Anodal IRE applications result in significantly less gas formation than both cathodal IRE applications and RF applications. This finding is supported by TOE observations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。