Misdiagnosis of narcolepsy caused by a false-positive orexin-A/hypocretin-1 enzyme immune assay

食欲素A/下丘脑泌素1酶免疫测定假阳性导致嗜睡症误诊

阅读:1

Abstract

The diagnosis of narcolepsy is based on clinical history, sleep studies, and, in some cases, cerebrospinal fluid orexin-A/hypocretin-1 measurement. The gold standard for orexin measurement is the radioimmunoassay but other commercial kits are also available, such as the enzyme immune assay (EIA). The specificity of orexin EIA in humans is unknown. We report four cases where orexin levels were measured by EIA and resulted in false positives and the misdiagnosis of narcolepsy. Therefore, orexin EIA measurement should be strongly discouraged in a clinical setting. CITATION: Sarkanen T, Sved G, Juujärvi M, Alakuijala A, Partinen M. Misdiagnosis of narcolepsy caused by a false-positive orexin-A/hypocretin-1 enzyme immune assay. J Clin Sleep Med. 2022;18(8):2075-2078.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。