Home monitoring for clinically suspected obstructive sleep apnea in pregnancy

孕期疑似阻塞性睡眠呼吸暂停的家庭监测

阅读:1

Abstract

STUDY OBJECTIVES: To determine if a home sleep apnea test (HSAT) using a type III portable monitor (PM), Nox-T3 (Nox Medical, Inc., Reykjavik, Iceland), detects obstructive sleep apnea in pregnant women. METHODS: Ninety-two pregnant women (34.5 ± 4.3 years; gestational age 25.4 ± 8.9 weeks; body mass index 29.9 ± 4.7 kg/m(2)) with suspected obstructive sleep apnea underwent HSAT with the Nox-T3 PM followed by overnight polysomnography (PSG) and PM recording simultaneously in the laboratory within 1 week. PMs were scored automatically and manually using a 3% criteria and compared with PSGs scored by following guidelines. RESULTS: Apnea-hypopnea indexes were 8.56 ± 10.42, 8.19 ± 13.79, and 8.71 ± 14.19 events/h on HSAT, in-laboratory PM recording, and PSG (P = .955), respectively. Bland-Altman analysis of the apnea-hypopnea index on PSG vs HSAT showed a mean difference (95% confidence interval) of -0.15 (-1.83, 1.53); limits of agreement (± 2 SD) were -16.26 to 16.56 events/h. Based on a threshold apnea-hypopnea index ≥ 5 events/h, HSAT had 91% sensitivity, 85% specificity, 84% positive-predictive value, and 92% negative-predictive value compared with PSG. When comparing the simultaneous recordings, closer agreement was observed. Automated vs manual analysis of PM showed no significant difference. CONCLUSIONS: A type III PM had an acceptable failure rate and high diagnostic performance operating as a reasonable alternative for in-laboratory PSG in pregnant women. CITATION: Wang J, Zhang C, Xu L, et al. Home monitoring for clinically suspected obstructive sleep apnea in pregnancy. J Clin Sleep Med. 2023;19(11):1951-1960.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。