The effect of design parameters of interspinous implants on kinematics and load bearing: an in vitro study

棘突间植入物设计参数对运动学和承载能力的影响:一项体外研究

阅读:1

Abstract

INTRODUCTION: A number of concepts with controversy approaches are currently discussed for interspinous stabilization (IPS). However, comparative biomechanical studies among the different systems are rare. Nevertheless, it remains unclear which biomechanical characteristics are influenced by different design features of these implants, such as implant stiffness or an additional tension band. Therefore, the aim of the present study was to compare different interspinous implants to investigate the biomechanical impact of IPS implant design on intersegmental kinematics, such as range of motion, neutral zone, center of rotation (COR), as well as load transfer like intradiscal pressure (IDP), to gain additional experience for clinical indications and limitations. MATERIAL AND METHOD: Twelve human lumbar spine specimens were tested in a spine loading apparatus. In vitro flexibility testing was performed by applying pure bending moments of 7.5 Nm without and with additional preload of 400 N in the three principal motion planes. Four interspinous implants, Coflex "COF" (Paradigm Spine, Germany), Wallis "WAL" (Abbott Laboratories, France), DIAM "DIA" (Sofamor Danek, France) and InterActiv (Aesculap AG, Germany) with two treatment options (without dorsal tensioning "IAO" and with dorsal tensioning "IAM") were consecutively tested in comparison to the native situation "NAT" and to a defect situation "DEF" of the functional spinal unit. The tested IPS devices are comprised of a compression stiffness range of 133 to 1,674 N/mm and a tensile stiffness range of 0-39 N/mm. Range of motion, neutral zone, center of rotation and intradiscal pressure were analyzed for all instrumentation steps and load cases. CONCLUSION: For the IPS, we found a correlation between compression stiffness and stabilization in extension. Here, the system with the lowest stiffness, DIA, displayed nearly no stabilization of the treated segment, whereas the system with the highest stiffness, WAL and COF, was most pronounced. This applies also for the correlation between device stiffness and IDP. In flexion only the degree of stabilization is in correlation with the tensile stiffness, whereas the IDP stays constant and is not affected by the different tensile stiffness. IPS is not able to stabilize in the frontal and transversal plane. Furthermore IPS does not substantially alter the location of the COR.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。