Comparison of revision strategies for failed C2-posterior cervical pedicle screws: a biomechanical study

C2后路颈椎椎弓根螺钉失败后翻修策略的比较:一项生物力学研究

阅读:1

Abstract

STUDY PURPOSE: With increasing usage within challenging biomechanical constructs, failures of C2 posterior cervical pedicle screws (C2-pCPSs) will occur. The purpose of the study was therefore to investigate the biomechanical characteristics of two revision techniques after the failure of C2-pCPSs. MATERIALS AND METHODS: Twelve human C2 vertebrae were tested in vitro in a biomechanical study to compare two strategies for revision screws after failure of C2-pCPSs. C2 pedicles were instrumented using unicortical 3.5-mm CPS bilaterally (Synapse/Synthes, Switzerland). Insertion accuracy was verified by fluoroscopy. C2 vertebrae were potted and fixed in an electromechanical testing machine with the screw axis coaxial to the pullout direction. Pullout testing was conducted with load and displacement data taken continuously. The peak load to failure was measured in newtons (N) and is reported as the pullout resistance (POR). After pullout, two revision strategies were tested in each vertebra. In Group-1, revision was performed with 4.0-mm C2-pCPSs. In Group-2, revision was performed with C2-pedicle bone-plastic combined with the use of a 4-mm C2-pCPSs. For the statistical analysis, the POR between screws was compared using absolute values (N) and the POR of the revision techniques normalized to that of the primary procedures (%). RESULTS: The POR of primary 3.5-mm CPSs was 1,140.5 ± 539.6 N for Group-1 and 1,007.7 ± 362.5 N for Group-2; the difference was not significant. In the revision setting, the POR in Group-1 was 705.8 ± 449.1 N, representing a reduction of 38.1 ± 32.9 % compared with that of primary screw fixation. For Group-2, the POR was 875.3 ± 367.9 N, representing a reduction of 13.1 ± 23.4 %. A statistical analysis showed a significantly higher POR for Group-2 compared with Group-1 (p = 0.02). Although the statistics showed a significantly reduced POR for both revision strategies compared with primary fixation (p < 0.001/p = 0.001), the loss of POR (in %) in Group-1 was significantly higher compared with the loss in Group-2 (p = 0.04). CONCLUSIONS: Using a larger-diameter screw combined with the application of a pedicle bone-plastic, the POR can be significantly increased compared with the use of only an increased screw diameter.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。