Validation of a novel spinal posture monitor: comparison with digital videofluoroscopy

新型脊柱姿势监测仪的验证:与数字视频透视法的比较

阅读:1

Abstract

PURPOSE: A novel, minimally invasive posture monitor which can monitor lumbar postures outside the laboratory has demonstrated excellent reliability, as well as concurrent validity compared to a surface marker-based motion analysis system. However, it is unclear if this device reflects underlying vertebral motion. METHODS: Twelve participants performed full range sagittal plane lumbo-pelvic movements during sitting and standing. Their posture was measured simultaneously using both this device (BodyGuard™) and digital videofluoroscopy. RESULTS: Strong correlations were observed between the two methods (all r (s) > 0.88). Similarly, the coefficients of determination were high (all r (2) > 0.78). The maximum mean difference between the measures was located in the mid-range of motion and was approximately 3.4° in sitting and 3.9° in standing. CONCLUSION: The BodyGuard™ appears to be a valid method for analysing vertebral motion in the sagittal plane and is a promising tool for long-term monitoring of spinal postures in laboratory and clinical settings in people with low back pain.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。