Biomechanical effect of different lumbar interspinous implants on flexibility and intradiscal pressure

不同腰椎棘突间植入物对柔韧性和椎间盘内压的生物力学影响

阅读:1

Abstract

Interspinous implants are used to treat lumbar spinal stenosis or facet joint arthritis. The aims of implanting interspinous devices are to unload the facet joints, restore foraminal height and provide stability especially in extension but still allow motion. The aim of this in vitro study was to compare four different interspinous implants--Colfex, Wallis, Diam and X-Stop--in terms of their three-dimensional flexibility and the intradiscal pressure. Twenty-four human lumbar spine specimens were divided into four equal groups and tested with pure moments in flexion/extension, lateral bending and axial rotation: (1) intact, (2) defect, (3) after implantation. Range of motion and the intradiscal pressure were determined. In each implant-group the defect caused an increase in range of motion by about 8% in lateral bending to 18% in axial rotation. Implantation had similar effects with all four implants. In extension, Coflex, Wallis, Diam, and X-Stop all overcompensated the instability caused by the defect and allowed about 50% of the range of motion of the intact state. In contrast, in flexion, lateral bending and axial rotation the values of the range of motion stayed about the values of the defect state. Similarly the intradiscal pressure after implantation was similar to that of the intact specimens in flexion, lateral bending and axial rotation but much smaller during extension. All tested interspinous implants had a similar effect on the flexibility: they strongly stabilized and reduced the intradiscal pressure in extension, but had almost no effect in flexion, lateral bending and axial rotation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。