Cervical spine disc prosthesis: radiographic, biomechanical and morphological post mortal findings 12 weeks after implantation. A retrieval example

颈椎间盘假体植入后12周的放射学、生物力学和形态学尸检结果。一个取出实例

阅读:2

Abstract

There is a gap between in vitro and clinical studies concerning performance of spinal disc prosthesis. Retrieval studies may help to bridge this gap by providing more detailed information about motion characteristics, wear properties and osseous integration. Here, we report on the radiographic, mechanical, histological properties of a cervical spine segment treated with a cervical spine disc prosthesis (Prodisc C, Synthes Spine, Paoli, USA) for 3 months. A 48-year-old male received the device due to symptomatic degenerative disc disease within C5-C6. The patient recovered completely from his symptoms. Twelve weeks later, he died from a subarachnoid hemorrhage. During routine autopsy, C3-T1 was removed with all attached muscles and ligaments and subjected to plain X-rays and computed tomography, three dimensional flexibility tests, shear test as well as histological and electronic microscopic investigations. We detected radiolucencies mainly at the cranial interface between bone and implant. The flexibility of the segment under pure bending moments of +/-2.5 Nm applied in flexion/extension, axial rotation and lateral bending was preserved, with, however, reduced lateral bending and enlarged neutral zone compared to the adjacent segments C4-C5, and C6-C7. Stepwise increase of loading in flexion/extension up to +/-9.5 Nm did not result in segmental destruction. A postero-anterior force of 146 N was necessary to detach the lower half of the prosthesis from the vertebra. At the polyethylene (PE) core, signs of wear were observed compared to an unused core using electronic microscopy. Metal and PE debris without signs of severe inflammatory reaction was found within the surrounding soft tissue shell of the segment. A thin layer of soft connective tissue covered the major part of the implant endplate. Despite the limits of such a case report, the results show: that such implants are able to preserve at least a certain degree of segmental flexibility, that direct bone implant contact is probably rare, and that debris may be found after 12 weeks.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。