Electrical counting and sizing of mammalian cells in suspension

悬浮状态下哺乳动物细胞的电计数和大小测定

阅读:1

Abstract

A recently developed method of determining the number and size of particles suspended in a conducting solution is to pump the suspension through a small orifice having an immersed electrode on each side to supply electrical current. The current changes due to the passage of particles of resistivity different from that of the solution. Theoretical expressions are developed which relate the current change caused by such particles to their volume and shape. It is found that most biological cells may be treated as dielectric particles whose capacitive effects are negligible. Electrolytic tank measurements on models confirm the theoretical development, and electric field plots of model orifices are used to predict the observed pulse shapes. An equivalent circuit of the orifice-electrode system is analyzed and shows that the current pulse may be made conductivity-independent when observed with a zero input impedance amplifier.