Modeling Intracochlear Magnetic Stimulation: A Finite-Element Analysis

耳蜗内磁刺激建模:有限元分析

阅读:2

Abstract

This study models induced electric fields, and their gradient, produced by pulsatile current stimulation of submillimeter inductors for cochlear implantation. Using finite-element analysis, the lower chamber of the cochlea, scala tympani, is modeled as a cylindrical structure filled with perilymph bounded by tissue, bone, and cochlear neural elements. Single inductors as well as an array of inductors are modeled. The coil strength (~100 nH) and excitation parameters (peak current of 1-5 A, voltages of 16-20 V) are based on a formative feasibility study conducted by our group. In that study, intracochlear micromagnetic stimulation achieved auditory activation as measured through the auditory brainstem response in a feline model. With respect to the finite element simulations, axial symmetry of the inductor geometry is exploited to improve computation time. It is verified that the inductor coil orientation greatly affects the strength of the induced electric field and thereby the ability to affect the transmembrane potential of nearby neural elements. Furthermore, upon comparing an array of micro-inductors with a typical multi-site electrode array, magnetically excited arrays retain greater focus in terms of the gradient of induced electric fields. Once combined with further in vivo analysis, this modeling study may enable further exploration of the mechanism of magnetically induced, and focused neural stimulation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。