Performance of continuous glucose monitoring systems (CGMs) during commercial flights in T1DM: A within-subject comparative pilot study

1型糖尿病患者在商业航班上使用连续血糖监测系统(CGM)的性能:一项受试者自身对照的试点研究

阅读:1

Abstract

AIMS: Continuous glucose monitoring (CGM) systems are increasingly used, but their accuracy under flight-conditions remains insufficiently studied. We assessed sensor accuracy of four CGMs during in-flight (mid-haul) and ground conditions. MATERIALS AND METHODS: Participants (n = 20) wore four sensors (Abbott Libre 3, Dexcom G7, Medtronic Simplera, and Sinocare iCan i3) in parallel, inserted 1 day prior to experiment. Capillary reference measurements were performed every 30 min. Accuracy evaluation included: proportion of CGM values within 10% of comparator glucose levels >100 mg/dL or within 10 mg/days of comparator levels ≤100 mg/dL (%10/10), along with %20/20 agreement rates, mean absolute relative difference (MARD), Consensus error grid. RESULTS: Twenty participants with type 1 diabetes (age: 37.9 ± 12.8 years, HbA1c: 52.3 ± 4.9 mmol/mol, diabetes duration 19.25 ± 10.6 years) were included in the analysis. Data from 297 sensors with 3473 matched pairs (G7/Libre3/Simplera/i3) were analysed. The overall MARDs were: 9.5% ± 3.3% (G7), 9.6% ± 3.8% (Libre3), 19.9% ± 6.2% (Simplera) and 15.7% ± 6.6% (i3). Overall %20/20 and %10/10 agreement rates were 90.6%/70.2% (G7), 90.8%/66.8% (Libre3), 63.6%/37.8% (Simplera) and 72.5%/42.2% (i3) respectively. The proportion of values in Zones A and B of the Consensus error grid was 98.9% (G7), 99.0% (Libre 3), 94.5% (Simplera) and 97.5% (i3). Prandial state and flight phase had limited influence on the relative performance ranking of the devices. CONCLUSIONS: G7 and Libre3 showed consistently higher accuracy and clinical reliability across various phases of flight and ground conditions. Simplera and i3 exhibited greater variability in accuracy. These findings suggest that while modern CGMs could support diabetes management in aviation, individual system performance may differ under specific operational conditions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。