Technical validation of an autoantibody test for lung cancer

肺癌自身抗体检测的技术验证

阅读:1

Abstract

BACKGROUND: Publications on autoantibodies to tumour-associated antigens (TAAs) have failed to show either calibration or reproducibility data. The validation of a panel of six TAAs to which autoantibodies have been described is reported here. MATERIALS AND METHODS: Three separate groups of patients with newly diagnosed lung cancer were identified, along with control individuals, and their samples used to validate an enzyme-linked immunosorbant assay. Precision, linearity, assay reproducibility and antigen batch reproducibility were all assessed. RESULTS: For between-replicate error, samples with higher signals gave coefficients of variation (CVs) in the range 7%-15%. CVs for between-plate variation were only 1%-2% higher. For between-run error, CVs were in the range 15%-28%. In linearity studies, the slope was close to 1.0 and correlation coefficient values were generally >0.8. The sensitivity and specificity of individual batches of antigen varied slightly between groups of patients; however, the sensitivity and specificity of the panel of antigens as a whole remained constant. The validity of the calibration system was demonstrated. CONCLUSIONS: A calibrated six-panel assay of TAAs has been validated for identifying nearly 40% of primary lung cancers via a peripheral blood test. Levels of reproducibility, precision and linearity would be acceptable for an assay used in a regulated clinical setting.