Dynamic assessment of mitral annular force profile in an ovine model

绵羊模型中二尖瓣环力曲线的动态评估

阅读:1

Abstract

BACKGROUND: Limited knowledge exists regarding the forces that act on devices implanted in the mitral annulus. Determining the peak magnitudes, directions, rates, variation throughout the cardiac cycle, and change with left ventricular pressure (LVP) will aid in device development and evaluation. METHODS: Novel transducers with the ability to measure forces in the septal-lateral and transverse directions were implanted in six healthy ovine subjects. Forces were measured for cardiac cycles reaching a peak LVP of 90, 125, 150, 175, and 200 mm Hg. RESULTS: The septal-lateral force was observed to significantly increase from 3.9 ± 0.8 N (90) to 5.2 ± 1.0 N (125) p < 0.001, 5.9 ± 0.9 N (150) p < 0.001, 6.4 ± 1.2 N (175) p < 0.001, and 6.7 ± 1.5 N (200 mm Hg) p < 0.001. Similarly, the transverse force was seen to increase from 2.6 ± 0.6 N (90) to 3.8 ± 1.0 N (125) p < 0.01, 4.6 ± 1.3 N (150) p < 0.001, 4.3 ± 1.2 N (175) p < 0.001, and 3.5 ± 0.7 N (200 mm Hg) p < 0.05. In comparison, the septal-lateral force was significantly greater than the transverse force at 90 (p < 0.05), 125 (p < 0.05), 175 (p < 0.001), and 200 mm Hg (p < 0.0005). CONCLUSIONS: Annular forces and their variations with LVP through the cardiac cycle are described. The results demonstrate differences in force magnitude and rate for increasing levels of LVP between the septal-lateral and transverse directions. These directional differences have strong implications in the development of future mitral devices.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。