Re-reading of OraQuick HIV-1/2 rapid antibody test results: quality assurance implications for HIV self-testing programmes

重新解读OraQuick HIV-1/2快速抗体检测结果:对HIV自检项目的质量保证启示

阅读:1

Abstract

INTRODUCTION: Scale-up of HIV self-testing (HIVST) will play a key role in meeting the United Nation's 90-90-90 targets. Delayed re-reading of used HIVST devices has been used by early implementation studies to validate the performance of self-test kits and to estimate HIV positivity among self-testers. We investigated the stability of results on used devices under controlled conditions to assess its potential as a quality assurance approach for HIVST scale-up. METHODS: 444 OraQuick(®) HIV-1/2 rapid antibody tests were conducted using commercial plasma from two HIV-positive donors and HIV-negative plasma (high-reactive n = 148, weak-reactive n = 148 and non-reactive n = 148) and incubated them for six months under four conditions (combinations of high and low temperatures and humidity). Devices were re-read daily for one week, weekly for one subsequent month and then once a month by independent readers unaware of the previous results. We used multistage transition models to investigate rates of change in device results, and between storage conditions. RESULTS AND DISCUSSION: There was a high incidence of device instability. Forty-three (29%) of 148 initially non-reactive results became false weak-reactive results. These changes were observed across all incubation conditions, the earliest on Day 4 (n = 9 kits). No initially HIV-reactive results changed to a non-reactive result. There were no significant associations between storage conditions and hazard of results transition. We observed substantial statistical agreement between independent re-readers over time (agreement range: 0.74 to 0.96). CONCLUSIONS: Delayed re-reading of used OraQuick(®) HIV-1/2 rapid antibody tests is not currently a valid methodological approach to quality assurance and monitoring as we observed a high incidence (29%) of true non-reactive tests changing to false weak-reactive and therefore its use may overestimate true HIV positivity.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。