Determination of MIC Quality Control Ranges for the Novel Gyrase Inhibitor Zoliflodacin

新型促旋酶抑制剂佐利氟达星的 MIC 质量控制范围的确定

阅读:9
作者:Alita A Miller, Maria M Traczewski, Michael D Huband, Patricia A Bradford, John P Mueller

Abstract

This report describes the results of two different, multilaboratory quality control (QC) studies that were used to establish QC ranges for the novel gyrase inhibitor zoliflodacin against the ATCC strains recommended by the Clinical and Laboratory Standards Institute (CLSI). Following the completion of an eight-laboratory, CLSI document M23-defined tier 2 study, the agar dilution MIC QC range for zoliflodacin against the Neisseria gonorrhoeae QC strain ATCC 49226 was defined as 0.06 to 0.5 μg/ml and was approved by the CLSI Subcommittee on Antimicrobial Susceptibility Testing. This QC range will be used for in vitro susceptibility testing of zoliflodacin during phase 3 human clinical trials and surveillance studies, and eventually it will be implemented in clinical labs. In a separate study, broth microdilution MIC quality control ranges for zoliflodacin against additional QC strains were determined to be 0.12 to 0.5 μg/ml for Staphylococcus aureus ATCC 29213, 0.25 to 2 μg/ml for Enterococcus faecalis ATCC 29212, 1 to 4 μg/ml for Escherichia coli ATCC 25922, 0.12 to 0.5 μg/ml for Streptococcus pneumoniae ATCC 49619, and 0.12 to 1 μg/ml for Haemophilus influenzae ATCC 49247. These MIC QC ranges were also approved by CLSI for use in future in vitro susceptibility testing studies against organisms other than N. gonorrhoeae.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。