An LC-MS-based designated comparison method with similar performance to the Lp(a) reference measurement procedure to guide molar Lp(a) standardization

基于 LC-MS 的指定比较方法,其性能与 Lp(a) 参考测量程序相似,用于指导摩尔 Lp(a) 标准化

阅读:9
作者:Nina M Diederiks #, L Renee Ruhaak #, Fred P H T M Romijn, Mervin M Pieterse, Nico P M Smit, Christa M Cobbaert

Background

The 2022 consensus statement of the European Atherosclerosis Society (EAS) on lipoprotein(a) (Lp(a)) recognizes the role of Lp(a) as a relevant genetically determined risk factor and recommends its measurement at least once in an individual's lifetime. It also strongly urges that Lp(a) test

Conclusions

The semi-automated, higher throughput, LC-MS-based method for Lp(a) meets the predefined analytical performance specifications and allowable MU and is hence applicable as a higher order Designated Comparison Method, which is ideally suited to guide IVD manufacturers in the transition from Lp(a) mass to molar units.

Methods

To assess whether the DCM provides equivalent

Results

The major difference in the design of the DCM for apo(a) is the use of only one enzymatic digestion step. The analytical performance of the DCM and RMP for apo(a) is highly similar. In a direct method comparison, equivalent results were obtained with a median regression slope 0.997 of and a median bias of - 0.2 nmol/L (- 0.2%); the intermediate imprecision of the test results was within total allowable error (TEa) (CVa of 10.2% at 90 nmol/L). Conclusions: The semi-automated, higher throughput, LC-MS-based method for Lp(a) meets the predefined analytical performance specifications and allowable MU and is hence applicable as a higher order Designated Comparison Method, which is ideally suited to guide IVD manufacturers in the transition from Lp(a) mass to molar units.