Conversion methods for modified Jaffe reaction assays of serum creatinine

血清肌酐改良Jaffe反应测定的转换方法

阅读:1

Abstract

BACKGROUND: Modifications in Jaffe serum creatinine (sCr) assays question the suitability of the results for direct comparison. METHODS: sCr in adult in-patients was routinely measured either by SRM 909-standardized/noncompensated (method A) or isotope dilution mass spectrometry traceable/compensated method (reference). We converted values by method A into values by the reference using a formula provided by the manufacturer [Beckman Coulter (BC)] and traditional equating methods. RESULTS: The BC-based conversion and linear equating resulted in underestimated sCr values, whereas equipercentile equating (EE) provided sCr with not significantly different distribution from the reference values. Proportions of patients with renal impairment did not differ between the reference and EE-converted sCr, as opposed to BC-recalculated values. Three percent of patients were classified into better renal function category when applying BC versus EE conversion. CONCLUSIONS: Equipercentile equation was a more accurate method for recalculation of sCr obtained from different Jaffe reaction assays than the linear equating or the BC linear formula. This study emphasizes the importance of the derivation sample specificity when applying research results to other real-world populations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。