Fracture of cobalt chrome, fully-coat beaded femoral revision long stem, a clinical outcomes study

钴铬合金全涂层珠状股骨翻修长柄假体断裂的临床结果研究

阅读:1

Abstract

BACKGROUND: Femoral bone loss during revision total hip arthroplasty poses a challenging problem. Bypass fixation over the diaphysis has achieved clinical success in cases of proximal femoral bone loss. Fracture of cementless, fully bead-coated femoral stem is an uncommon complication. The purpose of this study is to analyze the patients with and without fracture stem and find out the possible risk factors. METHODS: From 2006 to 2012, a total of 251 revision long stems (Zimmer, Warsaw, IN) were implanted. In the same period, 17 broken stems that underwent treatment were included for analysis. Patients' demographic data, pattern of femoral bone loss, stem size, medial calcar support in the proximal region of the stem, and the timing of stem breakage were collected and analyzed. RESULTS: The stem size in patients with a broken stem was smaller (p < 0.001), and medial calcar defect was 12.4% and 100% (p < 0.001), respectively. The bone defect was greater in broken group (p = 0.024). The mean duration between revision surgery and stem breakage was 58.07 ± 36.98 months. Smaller stem size, greater bone defect, and inadequate medial calcar bone support were major risk factors for stem breakage. CONCLUSIONS: Bypass fixation in the distal diaphysis with a long stem prosthesis without adequate bone support over medial calcar area may cause stress concentration in the long stem and a fatigue fracture. Use of a smaller prosthesis is the major risk of stem broken. It is essential to repair the proximal femoral bone deficiency and implant selection for better metaphyseal engagement to prevent further stem complications. LEVEL OF EVIDENCE: Level III, case control study.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。