Clinical Evaluation of the Measurement Performance of the Philips Health Watch: A Within-Person Comparative Study

飞利浦健康手表测量性能的临床评估:一项受试者自身对照研究

阅读:1

Abstract

BACKGROUND: Physical inactivity is an important modifiable risk factor for chronic diseases. A new wrist-worn heart rate and activity monitor has been developed for unobtrusive data collection to aid prevention and management of lifestyle-related chronic diseases by means of behavioral change programs. OBJECTIVE: The objective of the study was to evaluate the performance of total energy expenditure and resting heart rate measures of the Philips health watch. Secondary objectives included the assessment of accuracy of other output parameters of the monitor: heart rate, respiration rate at rest, step count, and activity type recognition. METHODS: A within-person comparative study was performed to assess the performance of the health watch against (medical) reference measures. Participants executed a protocol including 15 minutes of rest and various activities of daily life. A two one-sided tests approach was adopted for testing equivalence. In addition, error metrics such as mean error and mean absolute percentage error (MAPE) were calculated. RESULTS: A total of 29 participants (14 males; mean age 41.2, SD 14.4, years; mean weight 77.2, SD 10.2, kg; mean height 1.8, SD 0.1, m; mean body mass index 25.1, SD 3.1, kg/m(2)) completed the 81-minute protocol. Their mean resting heart rate in beats per minute (bpm) was 64 (SD 7.3). With a mean error of -10 (SD 38.9) kcal and a MAPE of 10% (SD 8.7%), total energy expenditure estimation of the health watch was found to be within the 15% predefined equivalence margin in reference to a portable indirect calorimeter. Resting heart rate determined during a 15-minute rest protocol was found to be within a 10% equivalence margin in reference to a wearable electrocardiogram (ECG) monitor, with a mean deviation of 0 bpm and a maximum deviation of 3 bpm. Heart rate was within 10 bpm and 10% of the ECG monitor reference for 93% of the duration of the protocol. Step count estimates were on average 21 counts lower than a waist-mounted step counter over all walking activities combined, with a MAPE of 3.5% (SD 2.4%). Resting respiration rate was on average 0.7 (SD 1.1) breaths per minute lower than the reference measurement by the spirometer embedded in the indirect calorimeter during the 15-minute rest, resulting in a MAPE of 8.3% (SD 7.0%). Activity type recognition of walking, running, cycling, or other was overall 90% accurate in reference to the activities performed. CONCLUSIONS: The health watch can serve its medical purpose of measuring resting heart rate and total energy expenditure over time in an unobtrusive manner, thereby providing valuable data for the prevention and management of lifestyle-related chronic diseases. TRIAL REGISTRATION: Netherlands trial register NTR5552; http://www.trialregister.nl/trialreg/admin/rctview.asp?TC=5552 (Archived by WebCite at http://www.webcitation.org/6neYJgysl).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。