Are the long term results of a high-flex total knee replacement affected by the range of flexion?

高屈曲度全膝关节置换术的长期效果是否受屈曲范围的影响?

阅读:1

Abstract

PURPOSE: The purpose of this study was to examine the effect of deep flexion on the long-term durability of a high-flex ceramic implant in total knee arthroplasty. METHODS: Five hundred and five consecutive knees replaced with a Bi-Surface knee system (Kyocera Medical, Osaka, Japan) were divided into two groups according to the range of flexion by 135° postoperatively. Comparison of implant durability was made between the high- and low-flexion groups after a minimum ten year follow-up. RESULTS: With revision for any surgery as the end point, the survival rates at ten years were 95.5% and 96.2% in the high- and low-flexion group, respectively (p = 0.63). With revision for mechanical failure as the end point, survival rates were 98.7% and 98.5%, respectively (p = 0.94). CONCLUSION: Implant survival rate was similar for both groups. Deep flexion seemed not to affect long-term durability.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。