Aseptic stem loosening in primary THA: migration analysis of cemented and cementless fixation

初次全髋关节置换术中无菌性股骨柄松动:有骨水泥固定和无骨水泥固定的移位分析

阅读:1

Abstract

Early migration has reportedly been predictive for later implant failure. Using four different migration patterns, this study aimed to analyse migration behaviour of the two types of implant fixation--cemented and cementless--throughout the process of loosening. Migrational behaviour of 69 revised stems (49 cemented, 20 uncemented) was analysed retrospectively with EBRA-FCA (Einzel-Bild-Röntgen-Analyse, Femoral Component Analysis). Uncemented stems failed after early and late onset migration alike, while late migration was the predominant pattern in cemented stems. Mean prosthetic failure after early migration occurred 5.8 (+/-4.4) years postoperatively due to insufficient primary stability. Initially stable stems with late onset migration were revised after 12.4 (+/-4.5) years. Measurement of early migration was found to be a valuable tool to screen short-term and mid-term failure. In the long run the method's sensitivity decreased. Late onset migration, however, preceded long-term failure by a mean of three years.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。