Tension-band wiring of transverse fractures of patella. The effect of site of wire twists and orientation of stainless steel wire loop: a biomechanical investigation

髌骨横形骨折的张力带钢丝固定术:钢丝扭转位置和不锈钢丝环方向的影响:一项生物力学研究

阅读:1

Abstract

Tension-band wiring is commonly used to treat transverse patella fractures. The most common configuration has parallel Kirschner (K) wires and a stainless steel wire loop placed in a vertically oriented figure of eight. Early mobilisation is important but can cause detrimental displacement at the fracture site. A wooden model of a transversely fractured patella was used to compare different fixation constructs. The fracture was fixed using two K wires and a stainless steel wire loop. One or two wire twists were used, placed either arbitrarily or at corners of the loop. The loop was oriented either vertically or horizontally. Interfragmentary compression at the point of wire breakage and permanent displacement on cyclic loading was measured for each construct. Placement of the figure of eight in a horizontal orientation with two wire twists at the corner improved interfragmentary compression by 63% (p<0.05, Tukey post hoc test). On cyclic loading, all constructs with vertical figure of eight but none with a horizontal construct failed (p=0.01; Fisher's exact test). Permanent fracture displacement after cyclic loading was 67% lower with horizontal figure of eight constructs (p<0.05; t test). Placing wire twists at the corner and a horizontal placement of figure of eight improves stability of the construct.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。