Ten-year follow-up of an anatomical hydroxyapatite-coated total hip prosthesis

解剖型羟基磷灰石涂层全髋关节假体的十年随访

阅读:2

Abstract

We report our results after ten year follow-up of 107 consecutive ABG-I hip prostheses implanted between June 1990 and December 1992: Only 84 prostheses were still in the study after ten years, but only six patients had undergone surgical revision. We can consider our clinical outcomes as excellent, with a whole-implant survival rate greater than 96%, a mean Merle D'Aubigne and Postel score increasing from 7.97 before operation to 16.17 at ten year follow-up, and a personal subjective assessment as excellent or good in 82.14% of patients. However, radiographic outcomes are more worrying: around 90% of patients show a stress-shielding phenomenon and granulomatous lesions in the proximal femur, and more than 82% suffer polyethylene wear greater than one millimetre (mean 1.68 mm). We think that zirconia stem heads and hooded antiluxation PE inserts are determining factors in the process of PE wear and, secondarily, in cancellous bone resorption and bone osteolysis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。