New, inexpensive and simple 3D printable device for nephelometric and fluorimetric determination based on smartphone sensing

一种基于智能手机传感技术的新型、低成本、简易的3D打印浊度法和荧光法测定装置

阅读:2

Abstract

A new, inexpensive and easy to use 3D printable device was developed for nephelometric and fluorimetric determination. Its applicability was tested for the quantification of quinine in tonic drinks and sulfate in natural water with good analytical accuracy. In this way, sulfate determination was carried out by nephelometry using a red LED, while quinine was determined using a blue LED by fluorimetry. A smartphone camera was used to take the pictures and afterwards transform them into the RGB color space using the software ImageJ by a personal computer. The linear range was 2.0-50.0 mg L(-1) for sulfate with a LOD of 0.13 mg L(-1), and the corresponding quantification limit (LOQ) was 0.43 mg L(-1). The linear range for quinine was from 0.42 to 3.10 mg L(-1). The LOD and LOQ were 0.11 mg L(-1) and 0.38 mg L(-1), respectively.