Transjugular tricuspid valve-in-ring replacement using the novel system under three-dimensional printing guidance

在三维打印引导下,使用新型系统进行经颈静脉三尖瓣环置换术

阅读:1

Abstract

Recurrent tricuspid regurgitation after failed ring annuloplasty is associated with high mortality rates and unfavourable clinical outcomes. Transcatheter tricuspid valve-in-ring (TVIR) replacement offers an alternative to high-risk surgical reoperation, but challenges remain. We present a case of a 67-year-old male who underwent tricuspid annuloplasty 3 years ago and developed severe recurrent tricuspid regurgitation (regurgitation volume = 36 mL) 6 months ago. To avoid a high-risk thoracotomy, we finally performed TVIR replacement using the novel device, LuX-Valve Plus system, via the transjugular approach. Cardiovascular three-dimensional printing technology was utilized to investigate the distinct advantages offered by the new device in this specific intervention, as well as to simulate the procedure. Successful TVIR replacement resulted in significant symptomatic improvement. TVIR replacement with the LuX-Valve Plus system is feasible for failed annuloplasty ring cases. Patient-specific three-dimensional printed models enable personalized procedural planning, enhancing safety and precision.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。