An Integrated System for Fetal Scalp Visualization, Blood Collection and Analysis

用于胎儿头皮可视化、血液采集和分析的集成系统

阅读:2

Abstract

KEY MESSAGE: The new NB scope aids in better visualization of the scalp and blood collection and analysis at bed side. OBJECTIVE: Caesarean section rates and inherent complications are on the rise all over the world. One way to avoid a caesarean is to measure fetal scalp blood lactate levels. The methods available to visualize fetal scalp, obtain the blood sample and perform the blood test are separate, cumbersome and expensive, needing a certain level of expertise. We propose a device that incorporates all the steps of obtaining a fetal scalp blood lactate into one sleek, easy to use device. METHODS: The initial design, 3-D print and was tried on mannequin. After ethics committee approval, the prototype was experimented on patients in labour with singleton live fetus in cephalic presentation with no evidence of distress. RESULTS: There were (n = 9) patients recruited. There were (n = 5) primigravida and (n = 4) multigravida all of whom were in active labour. Parity did not seem to influence ease of instrumentation. Of the (n = 9) mothers (n = 2) had meconium-stained liquor and the rest (n = 7) had clear liquor, meconium-stained liquor did not affect visualization. The mean time taken to collect the sample was 184.11(± 33.04) seconds. CONCLUSION: The Neeraj-Bhaskar (NB) scope is an easy to use, affordable device that can be used time and again to decide on cases where emergency caesarean section can be avoided due to fetal distress.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。