A Large-Diameter Vascular Graft Replacing Animal-Derived Sealants With an Elastomeric Polymer

大直径血管移植物用弹性聚合物替代动物源密封剂

阅读:8
作者:Manon Van Hecke, Lucas Van Hoof, Magdalena Sikole, Hozan Mufty, Piet Claus, Peter Verbrugghe, John Ely, Geoffrey A Berg, Tania Roskams, Bart Meuris

Conclusions

The experimental graft has a similar safety and performance profile and largely comparable necropsy results, in comparison to a commonly used prosthetic vascular graft, with the experimental grafts eliciting a nonadherent external fibrous capsule as the major difference compared to the control grafts that were incorporated into the periadventitia. Survival, hemostatic sealing, and hematologic and radiologic results were comparable between the study groups.

Methods

The experimental vascular grafts as well as control gelatin sealed interposition grafts were inserted into the descending aorta of juvenile sheep. The grafts were assessed by time to hemostasis and blood loss during surgery and hematology and biochemistry panels at distinct time points. Magnetic resonance imaging (MRI) was performed at 3 and at 6 mo after surgery, after which the animals were euthanized and necropsies were carried out including macroscopic and microscopic examination of the grafts, anastomoses, and distal organs.

Results

All animals survived the study period. There was no perceivable difference in the surgical handling of the grafts. The median intraoperative blood loss was 27.5 mL (range 10.0-125.0 mL) in the experimental group and 50.0 mL (range 10.0-75.0 mL) in the control group. The median time to hemostasis was 5.0 min (range 2.0-16.0 min) minutes in the experimental group versus 6.0 min (range 4.0-6.0 min) in the control group. MRI showed normal flow and graft patency in both groups. Healing and perianastomotic endothelialization was similar in both groups. Conclusions: The experimental graft has a similar safety and performance profile and largely comparable necropsy results, in comparison to a commonly used prosthetic vascular graft, with the experimental grafts eliciting a nonadherent external fibrous capsule as the major difference compared to the control grafts that were incorporated into the periadventitia. Survival, hemostatic sealing, and hematologic and radiologic results were comparable between the study groups.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。