Test-retest variability and reference region-based quantification of (18)F-BCPP-EF for imaging mitochondrial complex I in the human brain

(18)F-BCPP-EF在人脑线粒体复合物I成像中的重测变异性和基于参考区域的定量分析

阅读:1

Abstract

Mitochondrial complex I (MC-I) is an essential regulator of brain bioenergetics and can be quantified in the brain using PET radioligand (18)F-BCPP-EF. Here we evaluate the test-retest reproducibility of (18)F-BCPP-EF in humans, and assess the use of a non-invasive quantification method (standardised uptake value ratio - SUVR). Thirty healthy volunteers had a 90-min dynamic (18)F-BCPP-EF scan with arterial blood sampling, five of which received a second scan to be included in the test-retest analysis. Time-activity curves (TAC) were analysed using multilinear analysis 1 (MA1) and the two-tissue compartment model (2TC) to estimate volumes of distribution (V(T)). Regional SUVR-1 values were calculated from the 70 to 90-min TAC data using the centrum semiovale as a pseudo reference region, and compared to kinetic analysis-derived outcome measures. The mean absolute test-retest variability of V(T) ranged from 12% to 18% across regions. Both DVR-1and SUVR-1 had improved test-retest variability in the range 2%-7%. SUVR-1 was highly correlated with DVR-1 (r(2) = 0.97, n = 30). In conclusion, (18)F-BCPP-EF has suitable test-retest reproducibility and can be used to quantify MC-I in clinical studies.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。