Interference of Asfotase Alfa in Immunoassays Employing Alkaline Phosphatase Technology

Asfotase Alfa 对采用碱性磷酸酶技术的免疫测定的干扰

阅读:2
作者:Isabelle Danielle Piec, Beatrice Tompkins, William Duncan Fraser

Background

Asfotase alfa (STRENSIQ®, Alexion Pharmaceuticals, Inc.) is the only approved treatment for patients with pediatric-onset hypophosphatasia, a disease caused by a mutation in the tissue-nonspecific alkaline phosphatase (TNSALP) gene. ALP is often used as signaling system in routine immunoassays. Because asfotase alfa contains the active site of the full ALP enzyme, it can catalyze the substrate as the antibody-conjugated ALP would within an assay. Therefore, its presence in a treated patient's sample may generate false positive or false negative

Conclusion

The presence of asfotase alfa must be taken into consideration when analyzing blood samples in treated patients to avoid any risk of misinterpretation of false positive/negative results. It is essential that assays be tested for this possible interference.

Methods

Asfotase alfa was added to samples at concentrations from 0.08-5 µg/mL and analysed on various immunoassays following manufacturer's instructions.

Results

Asfotase alfa was detected in all ALP assays but ALKP1 (RayBiotech). We observed no changes in normetanephrine and noradrenaline (IBL) at any asfotase alfa concentration. However, asfotase alfa notably interfered in an oxytocin (ENZO) assay in nonextracted samples. Extraction using a C18 column eliminated the interference. No interference was observed on automated analyzers using alternative detection system (COBAS fT4 and TSH; Advia Centaur FSH, fT4; Architect LH; FSH). Immulite 2000 fT4, TSH, testosterone and hCG (ALP-based) showed no interference. However, the presence of asfotase alfa resulted in a dose-dependent increase of Troponin I signal.

特别声明

1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。

2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。

3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。

4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。