A multi-modality approach for enhancing 4D flow magnetic resonance imaging via sparse representation

一种基于稀疏表示的增强4D流动磁共振成像的多模态方法

阅读:1

Abstract

This work evaluates and applies a multi-modality approach to enhance blood flow measurements and haemodynamic analysis with phase-contrast magnetic resonance imaging (4D flow MRI) in cerebral aneurysms (CAs). Using a library of high-resolution velocity fields from patient-specific computational fluid dynamic simulations and in vitro particle tracking velocimetry measurements, the flow field of 4D flow MRI data is reconstructed as the sparse representation of the library. The method was evaluated with synthetic 4D flow MRI data in two CAs. The reconstruction enhanced the spatial resolution and velocity accuracy of the synthetic MRI data, leading to reliable pressure and wall shear stress (WSS) evaluation. The method was applied on in vivo 4D flow MRI data acquired in the same CAs. The reconstruction increased the velocity and WSS by 6-13% and 39-61%, respectively, suggesting that the accuracy of these quantities was improved since the raw MRI data underestimated the velocity and WSS by 10-20% and 40-50%, respectively. The computed pressure fields from the reconstructed data were consistent with the observed flow structures. The results suggest that using the sparse representation flow reconstruction with in vivo 4D flow MRI enhances blood flow measurement and haemodynamic analysis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。