Clinical and analytical comparison of six Simoa assays for plasma P-tau isoforms P-tau181, P-tau217, and P-tau231

对血浆 P-tau 亚型 P-tau181、P-tau217 和 P-tau231 的六种 Simoa 检测进行临床和分析比较

阅读:11
作者:Sherif Bayoumy #, Inge M W Verberk #, Ben den Dulk, Zulaiga Hussainali, Marissa Zwan, Wiesje M van der Flier, Nicholas J Ashton, Henrik Zetterberg, Kaj Blennow, Jeroen Vanbrabant, Erik Stoops, Eugeen Vanmechelen, Jeffrey L Dage, Charlotte E Teunissen

Discussion

P-tau isoforms can be measured robustly by several novel high-sensitive Simoa assays.

Methods

We studied the analytical (sensitivity, precision, parallelism, dilution linearity, and recovery) and clinical (40 AD dementia patients, age 66±8years, 50%F; 40 age- and sex-matched controls) performance of the assays.

Results

All assays showed robust analytical performance, and particularly P-tau217 Eli Lilly; P-tau231 Gothenburg and all P-tau181 assays showed robust clinical performance to differentiate AD from controls, with AUCs 0.936-0.995 (P-tau231 ADx: AUC = 0.719). Results obtained with all P-tau181 assays, P-tau217 Eli Lilly assay, and P-tau231 Gothenburg assay strongly correlated (Spearman's rho > 0.86), while correlations with P-tau231 ADx results were moderate (rho < 0.65).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。