Comparison of cytotoxicity between 3D printable resins and heat-cure PMMA

3D 打印树脂与热固化 PMMA 的细胞毒性比较

阅读:9
作者:N Gopi Chander, Aashmika Mahajan

Aim

The aim of this study was to evaluate and compare the cytotoxicity of polyurethane and polyoxymethylene printable resins with conventional heat cure polymethyl methacrylate denture base resins.

Conclusion

Polyoxymethylene resin exhibited the minimum cytotoxic properties among the tested materials, followed by polymethylmethacrylate and polyurethane resin.

Methods

The study followed ISO-10993-5 guidelines. It comprised of three groups. Fifteen cuboidal samples measuring 10x10 × 10mm dimension were prepared for each group. The polymethylmethacrylate samples were fabricated using conventional denture processing techniques, while the polyoxymethylene samples were printed using fused deposition modeling and the polyurethane samples using stereolithography technique. Post fabrication the samples were evaluated for cytotoxicity using the MTT assay with the VERO cell line. The percentage of cell viability was calculated to determine the cytotoxic effects.

Results

Statistical analysis revealed a significant difference in the cell viability of the experimental groups (p ≤ 0.0001). The polyoxymethylene group showed the highest % cell viability (62.78 %), followed by the polymethylmethacrylate group (52.43 %), and the least was observed in the polyurethane-based resin group (46.47 %). The findings indicate polyoxymethylene group displayed least cytotoxicity, followed by polymethylmethacrylate, and polyurethane-based resin.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。