Assessment of the FreeStyle Libre 2 interstitial glucose monitor in hypo- and euglycemic cats

对低血糖和正常血糖猫使用FreeStyle Libre 2组织间液葡萄糖监测仪的评估

阅读:1

Abstract

BACKGROUND: Continuous glucose monitoring systems have been validated for eu- and hyperglycemic cats. The FreeStyle Libre 2 (FSL2) is sufficiently accurate in people during hypoglycemia to guide critical treatment decisions without confirmation of blood glucose concentration (BG). OBJECTIVES: Assess FSL2 accuracy in cats with hypoglycemia. ANIMALS: Nine healthy, purpose-bred cats. METHODS: Hyperinsulinemic-hypoglycemic clamps were performed by IV infusion of regular insulin (constant rate) and glucose (variable rate). Interstitial glucose concentration (IG), measured by FSL2, was compared to BG measured by AlphaTrak2. Data were analyzed for all paired measurements (n = 364) and separately during stable BG (≤1 mg/dL/min change over 10 minutes). Pearson's r test, Bland-Altman test, and Parkes Error Grid analysis respectively were used to determine correlation, bias, and clinical accuracy (P < .05 considered significant). RESULTS: Overall, BG and IG correlated strongly (r = 0.83, P < .0001) in stable glycemia and moderately at all rates of change (r = 0.69, P < .0001). Interstitial glucose concentration underestimated BG in euglycemia, but the BG-IG difference was progressively smaller as BG decreased (12.9 ± 12.2, 8.8 ± 11.2, -3.2 ± 7.4, and -7.8 ± 5.2 mg/dL in the ranges of 80-120 [n = 64], 60-79 [n = 29], 50-59 [n = 71], and 29-49 mg/dL [n = 53], respectively). CONCLUSIONS: Although IG underestimates BG throughout most of the hypo-euglycemic range, IG generally overestimates BG in marked hypoglycemia (<60 mg/dL). It is therefore imperative to evaluate FSL2 results in this critical range with caution.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。