The biomechanical study of the influence to the forefoot plantar pressure of the first tarsometatarsal joint fracture-dislocation fixed by three different implants

三种不同植入物固定第一跖跗关节骨折脱位对前足足底压力影响的生物力学研究

阅读:1

Abstract

OBJECTIVES: To study the influence of forefoot plantar pressure of the first tarsometatarsal joint fracture-dislocation by three different implants to provide experimental reference in selecting implants. METHODS: Eight fresh foot specimens were made into the models of the first tarsometatarsal joint fracture-dislocation, which were fixed with 3.5 mm cortical screw, 1/4 tubular plate and compressive staple in turn. After the loading of 600N, the changes of the plantar pressure in forefoot were measured by the method of the F-scan plantar pressure system. RESULTS: After first tarsometatarsal joint fracture-dislocation, the peak pressure under the first metatarsal head would decrease, while the pressure under the second metatarsal head would increase(P<0.05). When the first tarsometatarsal joint was fixed with screw or plate respectively; the peak pressure under the two metatarsal heads would tend to be normal. However, the staple fixation would show the statistical significant difference, although the peak pressure under the first and second metatarsal heads were recovered in some extent(P<0.05). CONCLUSIONS: After the first tarsometatarsal joint fracture-dislocation, the plantar pressure might be compensated partly by the adjacent metatarsal heads. While the first tarsometatarsal joint fracture-dislocation was fixed by screw or plate, the plantar pressure of the forefoot would return to the normal state. However, if the joint was fixed by the staple, it would still be difficult to return the plantar pressure to be normal.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。