Effects of error on dose of target region and organs at risk in treating nasopharynx cancer with intensity modulated radiation therapy

强度调制放射治疗治疗鼻咽癌时,误差对靶区和危及器官剂量的影响

阅读:1

Abstract

OBJECTIVE: To measure setup error of head and neck neoplasm in radiotherapy and discuss over effects of error on physical dose acting on target region and organs at risk of nasopharynx cancer (NPC) patients treated with intensity modulated radiation therapy (IMRT). METHODS: A total of 152 patients who developed head and neck neoplasm and received IMRT were randomly selected. Through comparing digital portal image and digital reconstruction image, we measured setup error, calculated expanding margin from clinical target volume (CTV) to planning target volume (PTV) and analyzed whether there was rules between setup error and treatment time. Additionally, 20 cases of NPC were selected. Three-dimensional error was simulated in planning system. Dose distribution was recalculated and a series of dose parameters of target volume and OAR were analyzed. RESULTS: Setup error in left-right, head-feet and ventral-dorsal direction was (-0.62±1.46) mm, (-0.41±1.54) mm and (-0.31±1.67) mm respectively. Regarding limit value, the maximum and minimum value in left-right direction, head-feet direction and ventral-dorsal direction was 2.70 mm and -6.00 mm; 3.00 mm and -5.00 mm, 5.00 mm and -7.50 mm. Expanding margin from CTV to PTV was 2.26 mm, 1.88 mm and 1.97 mm in left-right direction, head-feet direction and ventral-dorsal direction. CONCLUSION: During IMRT, only when setup error is controlled below 3 mm can sharply reduce the damage caused by radiation to normal tissue; therefore, quality security and control of electronic portal imaging device need (EPID) to be improved.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。