Usability and Acceptability of the Abatacept Pre-Filled Autoinjector for the Subcutaneous Treatment of Rheumatoid Arthritis

阿巴西普预充式自动注射器在皮下治疗类风湿性关节炎中的可用性和可接受性

阅读:1

Abstract

INTRODUCTION: The purpose of the present study was to determine whether the abatacept autoinjector can be used by the intended population without patterns of preventable use errors, and is acceptable when assessed against key user needs. METHODS: Two independently conducted simulated-use studies, with no active drug administered, quantified use errors and evaluated the abatacept autoinjector and competitor devices on key attributes (comfort, control, ease of use, confidence of dose) and overall acceptability. Autoinjector preference was also assessed. Participants were patients with rheumatoid arthritis, caregivers, and healthcare professionals (HCPs). Participants were informed that a new rheumatoid arthritis autoinjector was being tested but were blinded to the intended drug and sponsor identity. RESULTS: In the formative (pre-validation) study (n = 54), two high-priority use errors occurred, both of which resulted from protocol non-compliance rather than mental confusion or physical limitations. In the summative (validation) study (n = 99), one high-priority use error occurred; this was deemed a simulated-use study artifact as participant behavior was guided by actual experience associated with the feel of drug delivery into the skin rather than by protocol, so no mitigation steps were considered necessary. Across user groups, average scores were consistently high for the pre-defined key attributes. Overall acceptability scores (7-point scale) were significantly higher for the abatacept versus competitor autoinjectors-formative study: patients 6.7 vs 5.2 (P = 0.0001), caregivers 7.0 vs 4.6 (P = 0.0093), HCPs 6.8 vs 5.1 (P = 0.0020); summative study: patients 6.5 vs 5.9 (P = 0.0404), caregivers 6.8 vs 5.8 (P = 0.0047), HCPs 6.8 vs 5.1 (P = 0.0002). The abatacept autoinjector was preferred to competitor devices: patients 85.7% vs 14.3% (P = 0.00002), caregivers 84.2% vs 15.8% (P = 0.00443), HCPs 95.0% vs 5.0% (P = 0.00004). Positive experiences with the abatacept autoinjector were attributed to the rubberized grip, device size, visualization of dose progression, button ergonomics, and ease of use. CONCLUSION: The abatacept autoinjector demonstrated usability without patterns of preventable use errors, and with high acceptability ratings across all key attributes assessed. Preference over competitor autoinjectors was due to device ergonomics, visualization of dose progression, confidence of dose delivery, and overall ease of use. FUNDING: Bristol-Myers Squibb.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。