Magnetic nanosensor for detection and profiling of erythrocyte-derived microvesicles

用于检测和分析红细胞衍生微囊泡的磁性纳米传感器

阅读:1

Abstract

During the course of their lifespan, erythrocytes actively shed phospholipid-bound microvesicles (MVs). In stored blood, the number of these erythrocyte-derived MVs has been observed to increase over time, suggesting their potential value as a quality metric for blood products. The lack of sensitive, standardized MV assays, however, poses a significant barrier to implementing MV analyses into clinical settings. Here, we report on a new nanotechnology platform capable of rapid and sensitive MV detection in packed red blood cell (pRBC) units. A filter-assisted microfluidic device was designed to enrich MVs directly from pRBC units, and label them with target-specific magnetic nanoparticles. Subsequent detection using a miniaturized nuclear magnetic resonance system enabled accurate MV quantification as well as the detection of key molecular markers (CD44, CD47, CD55). When the developed platform was applied, MVs in stored blood units could also be monitored longitudinally. Our results showed that MV counts increase over time and, thus, could serve as an effective metric of blood aging. Furthermore, our studies found that MVs have the capacity to generate oxidative stress and consume nitric oxide. By advancing our understanding of MV biology, we expect that the developed platform will lead to improved blood product quality and transfusion safety.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。