Accuracy of point-of-care capillary blood sugar measurements in critically ill patients: An observational study

危重患者床旁毛细血管血糖测量的准确性:一项观察性研究

阅读:1

Abstract

BACKGROUND: Accurately monitoring blood glucose levels is vital for critically ill individuals. Point-of-care (POC) glucose meters are commonly used in local intensive care units (ICUs). This study aimed to assess the precision of POC glucose meter readings in critically ill individuals with specific evaluation in patients with and without shock against the reference standard of venous blood glucose measurements. METHODS: An observational study was done on adult patients admitted in the ICU at a teaching institution. Capillary blood samples were collected from the patient's fingertip using lancet device with aseptic measures. The sample was analyzed using the GlucoCare Sense Glucometer (RMD Mediaids Limited, Taiwan). At the same time, 2 ml of blood was drawn from the patient's peripheral veins and analyzed by glucose oxidase-peroxidase method as reference. RESULTS: POC glucose measurements averaged 140 ± 20.23 mg/dl, while laboratory values were recorded as 116.10 ± 17.13 mg/dl. The difference between the two methods was 24.34 ± 12.01 mg/dl. A strong correlation (r = 0.805) was found between capillary and laboratory blood glucose levels, indicating a significant association (P < 0.0001). Twenty-two (44%) patients were in shock during the study. The mean difference between laboratory and POC blood glucose levels was higher in patients with circulatory shock (36.82 ± 4.84 mg/dl) than those without shock (14.61 ± 4.49 mg/dl), P < 0.05. CONCLUSION: POC glucose meters may lead to underdetection of hypoglycemia in critically ill patients, as their values are higher than laboratory values. Moreover, the results showed that POC glucometers are inaccurate for monitoring glucose in hypotensive patients in shock. Standard venous glucose monitoring methods may be more appropriate for these patients.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。