Multi-modal pain measurements in infants

婴儿多模式疼痛测量

阅读:1

Abstract

A non-invasive integrated method was developed to measure neural and behavioural responses to peripheral sensory and noxious stimulation in human infants. The introduction of a novel event-detection interface allows synchronous recording of: (i) muscle and central nervous system activity with surface electromyography (EMG), scalp electroencephalography (EEG) and near-infrared spectroscopy (NIRS); (ii) behavioural responses with video-recording and (iii) autonomic responses (heart rate, oxygen saturation, respiratory rate and cardiovascular activity) with electrocardiography (ECG) and pulse oximetry. The system can detect noxious heel lance and touch stimuli with precision (33 μs and 624 μs respectively) and accuracy (523 μs and 256 μs) and has 100% sensitivity and specificity for both types of stimulation. Its ability to detect response latencies accurately was demonstrated by a shift in latency of the vertex potential of 20.7 ± 15.7 ms (n=6 infants), following touch of the heel and of the shoulder, reflecting the distance between the two sites. This integrated system has provided reliable and reproducible measurements of responses to sensory and noxious stimulation in human infants on more than 100 test occasions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。