Adjustable Cortical Fixation Device for Quadriceps Tendon Repair: A Cadaveric Biomechanical Study

用于股四头肌腱修复的可调节皮质固定装置:一项尸体生物力学研究

阅读:1

Abstract

BACKGROUND: Adjustable cortical fixation devices have demonstrated utility in orthopaedic applications, such as ankle syndesmosis repair. PURPOSE: To assess the cyclic gap formation of a quadriceps tendon repair technique using an adjustable cortical fixation device compared with repair with knotless suture anchors and suture tape, a modification of conventional suture anchor repair. STUDY DESIGN: Controlled laboratory study. METHODS: Eight fresh-frozen matched pairs of cadaveric knees were used. Specimens in each pair were randomized to undergo either modified suture anchor repair (control) or adjustable cortical fixation repair. The control repair was performed as previously described. The experimental repair was performed using 2 No. 2 FiberWire sutures placed into the quadriceps tendon in a running locked Krackow configuration and 2 adjustable loop devices passed through transosseous tunnels. The lagging strands of the devices were tensioned to seat the cortical fixation buttons at the inferior patellar pole and then tied to the free Krackow strands at the superior pole to complete the repair. The mean plastic gap (permanent tendon displacement that did not recover with cyclic extension) and mean maximum gap (peak displacement that occurred with cyclic knee flexion and partially recovered with extension) were evaluated during cyclic loading for 500 cycles of full knee extension to 90° of flexion. RESULTS: At all testing intervals, the mean plastic gap was significantly smaller for the cortical fixation group versus the suture anchor group (P < .02). Similarly, the mean maximum gap was significantly smaller for the cortical fixation specimens at all testing intervals (P < .01). After cyclic loading, the mean maximum gap was significantly smaller in the cortical fixation group (4.80 ± 1.56 mm) versus the suture anchor group (8.47 ± 1.47 mm; P = < .001). The mean plastic gap was also significantly smaller in the cortical fixation versus the suture anchor group (3.25 ± 1.10 mm vs 6.57 ± 1.62 mm, respectively; P = < .001). CONCLUSION: Quadriceps tendon repair using an adjustable cortical fixation device demonstrated superior biomechanical properties in cyclic displacement testing compared with repair using the suture anchor technique. CLINICAL RELEVANCE: These results suggest that an adjustable cortical fixation device is a biomechanically viable alternative for quadriceps tendon repair.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。