Fixed- Versus Adjustable-Loop Femoral Cortical Suspension Devices for Anterior Cruciate Ligament Reconstruction: A Systematic Review and Meta-analysis of Biomechanical Studies

用于前交叉韧带重建的固定环与可调节环股骨皮质悬吊装置的比较:生物力学研究的系统评价和荟萃分析

阅读:1

Abstract

BACKGROUND: Fixed- and adjustable-loop femoral cortical suspension devices are commonly used for femoral graft fixation during anterior cruciate ligament reconstruction (ACLR). PURPOSE: To compare the biomechanical results of fixed- versus adjustable-loop femoral cortical suspension devices in studies simulating ACLR with an isolated device and/or specimen setup using porcine femora and bovine flexor tendons. STUDY DESIGN: Systematic review. METHODS: Two independent reviewers searched PubMed, Embase, and the Cochrane Library databases to find studies comparing the biomechanical strength of fixed- and adjustable-loop cortical suspension devices for ACLR with isolated device and/or specimen setups using porcine femora and bovine flexor tendons. Studies that compared both devices with similar biomechanical methods were included. Data extracted included displacement during cyclic loading, ultimate load to failure, and mode of failure of the different cortical suspension devices for ACLR. RESULTS: Six studies were identified that met the inclusion criteria, including a total of 76 fixed-loop devices and 120 adjustable-loop devices. Load to failure was significantly different (P < .0001), with the strongest fixation device being the ToggleLoc with ZipLoop adjustable-loop device (1443.9 ± 512.3 N), compared with the Endobutton CL fixed-loop device (1312.9 ± 258.1 N; P = .04) and the TightRope RT adjustable-loop device (863.8 ± 64.7 N; P = .01). Cyclic displacement was significantly different, with Endobutton CL (3.7 ± 3.9 mm) showing the least displacement, followed by ToggleLoc with ZipLoop (4.9 ± 2.3 mm) and TightRope RT (7.7 ± 11.1 mm) (P < .0001). Mode of failure was statistically different between the 3 groups (P = .01), with suture failure accounting for 83.8% of TightRope RT devices, 69.4% of ToggleLoc with ZipLoop devices, and 60.3% of Endobutton CL devices. CONCLUSION: Current biomechanical data suggest that the ToggleLoc with ZipLoop device is the strongest fixation device at "time zero" in terms of ultimate load to mechanical failure. However, the Endobutton CL device demonstrated the least cyclic displacement, which may be a more clinically applicable measure of device superiority.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。