Comparison of CT-Guided Microwave Ablation of Liver Malignancies with and Without Intra-Arterial Catheter Placement for Contrast Administration

比较CT引导下微波消融治疗肝脏恶性肿瘤时是否使用动脉内导管进行造影剂注射的效果

阅读:1

Abstract

BACKGROUND: The aim of this study was to compare microwave ablation (MWA) with and without prior placement of an intra-arterial catheter for the purpose of application of contrast medium (CM). METHODS: 148 patients (45 female, 65.1 ± 14.9 years) with liver tumors who underwent CT-guided MWA were included. Of these, 25 patients had an IA catheter placed in the hepatic artery. RESULTS: 37 patients underwent planning imaging for MWA without CM. A total of 86 patients received a standard dose of 80 mL intravenous (IV) CM for the planning scans. The patients with an IA catheter (n = 25) received an IA application of 10 mL CM. A total of 29 patients received contrast-enhanced scans in the PV phase for control of needle positioning after IV application of a standard dose of 80 mL CM. In patients with an IA catheter, control of the needle position was performed by single-slice scans. IA CM application during the ablation enabled monitoring of the ablation zone. Over the entire intervention, patients with IA catheters received less CM as compared to patients without an IA catheter (39.1 ± 10.4 mL vs. 141 ± 39.69 mL; p < 0.001). CONCLUSIONS: IA catheter placement was associated with a significant decrease of the amount of CM during MWA and enabled monitoring of the ablation zone.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。