No-Reference-Based and Noise Level Evaluations of Cinematic Rendering in Bone Computed Tomography

骨骼计算机断层扫描中电影渲染的无参考和噪声水平评估

阅读:1

Abstract

Cinematic rendering (CR) is a new 3D post-processing technology widely used to produce bone computed tomography (CT) images. This study aimed to evaluate the performance quality of CR in bone CT images using blind quality and noise level evaluations. Bone CT images of the face, shoulder, lumbar spine, and wrist were acquired. Volume rendering (VR), which is widely used in the field of diagnostic medical imaging, was additionally set along with CR. A no-reference-based blind/referenceless image spatial quality evaluator (BRISQUE) and coefficient of variation (COV) were used to evaluate the overall quality of the acquired images. The average BRISQUE values derived from the four areas were 39.87 and 46.44 in CR and VR, respectively. The difference between the two values was approximately 1.16, and the difference between the resulting values increased, particularly in the bone CT image, where metal artifacts were observed. In addition, we confirmed that the COV value improved by 2.20 times on average when using CR compared to VR. This study proved that CR is useful in reconstructing bone CT 3D images and that various applications in the diagnostic medical field will be possible.