Left Ventricular Assist Device Pump Obstruction Reduces Native Heart Efficiency

左心室辅助装置泵阻塞会降低自身心脏的效率

阅读:1

Abstract

Obstruction of the LVAD flow path can occur when blood clots or tissue overgrowth form within the inflow cannula, pump body, or outflow graft, and it can lead to thrombus, embolism, and stroke. The goal of this study was to measure the impact of progressive pump inflow obstruction on the pressure and flow dynamics of the LVAD-supported heart using a mock circulatory loop. Pump obstruction (PO) was produced by progressively blocking a fraction of the LVAD inlet area. Pressures, flows, and the midplane velocity field of the LV were measured for three LVAD speeds and six PO levels. Pressure and flow decreased with PO, shifting more of the flow through the aortic valve such that the total flow decreased by 6-11% and decreased the efficiency of the work of the native heart up to 60%. PO restricts diastolic flow through the LVAD, which reduces mitral inflow and decreases the strength and energy of the intraventricular vortices. The changes in flow architecture produced by PO include flow stasis and increased shear, which predispose the system to thromboembolic risk. Analysis of the contributions to external work may enable early detection, which allows time for therapeutic intervention, reducing the likelihood of pump replacement and the risk of complications.