Comparison of the TONOVET Plus®, TonoVet®, and Tono-Pen Vet™ tonometers in normal cats and cats with glaucoma

正常猫和青光眼猫的 TONOVET Plus®、TonoVet® 和 Tono-Pen Vet™ 眼压计的比较

阅读:1

Abstract

OBJECTIVES: To determine the accuracy, precision, and clinical applicability of the ICare® TONOVET Plus (TVP) in cats. ANIMALS AND PROCEDURES: IOP readings obtained with the TVP were compared to values obtained concurrently with the original TONOVET (TV01) and Tono-Pen Vet™ (TP) in 12 normal cats (24 eyes) and 8 glaucomatous LTBP2-mutant cats (13 eyes) in vivo. Reproducibility of TVP readings was also assessed for three observers in the above cats. The anterior chambers of five different normal cat eyes were cannulated ex vivo. IOP was measured with the TVP, TV01, and TP at manometric IOPs ranging from 5 to 70 mmHg. Data were analyzed by linear regression, ANOVA and Bland-Altman plots. ANOVA was used to assess reproducibility of TVP readings obtained by different observers and an ANCOVA model controlled for variation of individual cats. p < .05 was considered significant. RESULTS: TVP values strongly correlated with TV01 values (y = 1.045x + 1.443, R(2)  = .9667). The TP significantly underestimated IOP relative to the TVP and TV01, particularly at high IOP. IOP values obtained by 1 observer were significantly higher (~1 mmHg average) compared to the other 2 observers via ANCOVA analysis (p = .0006479 and p = .0203). Relative to manometry, the TVP and TV01 were significantly more accurate (p < .0001) and precise (p < .0070) than the TP in ex vivo eyes. CONCLUSIONS: IOP readings obtained with the TVP and TV01 are broadly interchangeable between models and between observers, but subtle differences may be important in a research context. TP readings vastly underestimate high IOP in feline glaucoma.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。