Repeatability of Corneal Densitometry Measurements using a Scheimpflug Camera in Healthy Normal Corneas

使用 Scheimpflug 相机对健康正常角膜进行角膜密度测量的重复性

阅读:3

Abstract

PURPOSE: To determine the repeatability of corneal densitometry measured by the Scheimpflug imaging system. METHODS: This cross-sectional study was conducted on photorefractive keratectomy candidates. One eye of each participant underwent imaging using Pentacam HR three times, 10 min apart. The repeatability of densitometry measurements was evaluated in four concentric annuli around the corneal apex and in different corneal depths. The repeatability of the measurements was evaluated using the intraclass correlation coefficient (ICC), repeatability coefficient (RC), and coefficient of variation (CV). The difference of repeatability between layers and zones was tested by tolerance index (TI). RESULTS: Sixty eyes of sixty patients with a mean age of 27.76 ± 3.93 years were studied. Half of the participants were female (n = 30, 50%). ICC was above 0.9 in all corneal parts. The posterior layer and central zones showed the least variability of densitometry measurements considering the CV values. The RC was 2.06, 1.17, and 0.92 in anterior, central, and posterior layers, respectively. The RC was 0.88, 0.71, 1.51, and 4.56 in 0-2, 2-6, 6-10, and 10-12 mm circles, respectively. Only the reliability of densitometry in 10-12 mm annulus was statistically lower than the central zone (TI = 0.71). CONCLUSIONS: Corneal densitometry measurements provided by the Pentacam had good repeatability. The repeatability of densitometry measurements decreased from the center to the periphery (with an exception for 0-2 mm and 2-6 mm) and from the posterior to the anterior of the cornea. The reliability of the 10-12 mm zone was markedly less than other zones.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。