Safety and biocompatibility of a bionic eye: Imaging, intraocular pressure, and histology data

仿生眼的安全性和生物相容性:影像学、眼内压和组织学数据

阅读:1

Abstract

The data presented here are related and supplementary data to the research article "Implantation and long-term assessment of the stability and biocompatibility of a novel 98 channel suprachoroidal visual prosthesis in sheep" [1]. In Eggenberger et al., nine sheep of the Suffolk (N=2) and Dorper (N=7) breeds were implanted in the left eye with an electrically inactive, suprachoroidal retinal stimulator (Bionic Eye) for durations of up to 100 days. The surgical safety, implant stability and device biocompatibility were assessed. Intraocular pressure measurements, indirect and infrared ophthalmoscopy and optical coherence tomography were performed at fixed time points to evaluate the clinical effects of the surgery and device implantation. Post-mortem eye tissue collection and histology was performed to measure the effects of the intervention at the cellular level. The data, including a comprehensive collection of fundus, infrared, optical coherence tomography and histology images can be used as a reference for comparison with other research, for example, active retinal stimulators. Furthermore, these data can be used to evaluate the suitability of the sheep model, in particular Dorper sheep, for future research.