Abstract
This document provides guidelines for the terminology, methodology, and for the interpretation of data obtained from the use of skin (transcutaneous) po(2) and pco(2) electrodes. The transcutaneous technique has found special application in newborn infants. The causes of analytical bias with respect to arterial blood gas values, and imprecision obtained with transcutaneous pco(2) electrodes, are reviewed. Electrode temperatures above 44 degrees C should not be used routinely, and, at a measuring temperature of 44 degrees C, the measuring site should be changed at least every 4 h to avoid skin burns.