Abstract
BACKGROUND: Fractured catheter as a foreign body in situ is a rare complication after port catheter placement. We report a single center's experience on percutaneous transvenous retrieval of intravascular fractured port catheter and treatment techniques. METHODS: Patients undergoing percutaneous transvenous retrieval of intravascular fractured port catheter from Jan 2010 to Dec 2018 were retrospectively collected. A total of 10 patients (8 females and 2 males) were enrolled in this study. Procedures were performed within 1 day after diagnosis. Two methods of retrieval were considered, direct retrieval by gooseneck snare and guide wire as media to retrieve were used in the procedure. RESULTS: All the fractured catheters in 10 patients were successfully retrieval by 2 methods, direct retrieval by gooseneck snare(n = 6) and guide wire as media of retrieval(n = 4). The time interval between port catheter implantation and discovery of catheter fracture was 36.50 ± 42.99(ranged 1 to 146) days. The operation time was 24.10 ± 8.32(ranged 10 to 36) minutes. No immediate procedure related or 1 month follow-up complications occurred in all the 10 patients. CONCLUSION: Percutaneous transvenous retrieval of intravascular fractured port catheter is a simple and safe procedure, which maybe recommended as the first choice for patients with fractured port catheter in situ.